A linear array transducer

with an arterial exam type

is used to perform an ultrasound-guided insertion

of a radial artery catheter via a transverse approach.

The patient is in a supine position

with the arm abducted 45 degrees and externally rotated.

The wrist is extended by placing a small roll

beneath the wrist crease.

The transducer is placed transversely

just proximal to the wrist crease,

with the orientation marker directed to the patient's right.

The radial artery is superficial and seen as an anechoic,

round, pulsatile structure on the lateral aspect

of the wrist amid static, echodense material.

There are two radial veins that accompany the radial artery,

which can be compressed with gentle transducer pressure.

The radius bone can be seen as a hyperechoic structure

approximately one centimeter posterior to the radial artery.

Adjust the transducer so it is centered

over the radial artery.

With the needle positioned just superior to the transducer,

slowly insert the needle

next to the centerline on the transducer.

The needle will appear as a small bright dot

in the ultrasound image.

When the needle tip appears, slide the transducer

a short distance away until the tip is no longer visible.

Once again, advance the needle slowly

until it is visible in the ultrasound image.

Repeat the process to ensure that the needle tip

is advanced under direct ultrasound guidance.

Continue advancing the needle until the tip

is seen to indent and then puncture the radial artery.

Confirm that the needle tip lies

in the mid-portion of the artery

by rotating the probe 90 degrees

to get an in-plane longitudinal view

of the needle within the artery.

手順:超音波ガイド下橈骨動脈アクセス

動脈検査用のリニアアレイトランスデューサーを用いて、横方向アプローチによる橈骨動脈カテーテルの超音波ガイド下挿入を行います。